ANESTHESIA VENTILTOR MODEL:DAV80

Author:dolphinmed 2019-12-16 11:44:35 357 0 0

China Medical Anesthesia, Anesthesia Workstation, Anesthesia Gas Machine Manufacturer

Feature

Suitable for a range of anesthesia at adultpediatric

Comprehensive Safety&Alarm systemmake sure the operation accuracy and stability

8 inch TFT display with touch screenshuttle knob design, quick and efficient operation

Powerful ventilation mode meet all complex clinical requirement

Three levels alarm system, visual and sound alarm information.

The core components are imported from top international and original, and the industrial control chip is chosen

High precious sensor and valve technology make sure the pressure ,flow and volume monitor and control at high safety, stability and precision

Built-in backup battery provide the emergency power supply to the unit.

 

Specifications anesthesia ventialtor

Display:8 " TFT, with touch screen.

Driving gas:O2 or Air,Inlet pressure,280-600 kPa.Max flow ≤ 120 L / min.

Mode:VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT.

Optional: SIMV-PRVC, PRVC

Waveform:Pressure,Flow-rate,Volume

Loop:PV loop, PF loop,FV loop

Safety Pressure:Does not exceed 6 KPa.

Battery Type:Li-ion,work time more than 2 hours

Parameter Setting Range:

Tidal Volume:Range: 20 - 1500ml; 

Respiratory rate::1~100 bpm;

Inspiratory time:0.1 ~ 10.0 s;

Respiratory ratio: 9.9:1 to 1:9.9;

Percentage of inspiratory pause:0 to 50%;

PEEP:OFF, 3 ~ 20 cmH2O

Pressure Support:0 ~ 60 cmH2O

Pressure Control:5 ~ 60 cmH2O

Flow trigger: 0.5 ~ 20L/min

Pressure Trigger: -1 ~ -20 cmH2O

PSV Insp Termination Level:25%

Monitoring Parameter

Inspiratory tidal volume,Expiratory tidal volume,Minute ventilation,Spontaneous minute ventilation,Respiratory rate,Spontaneous breathing frequency,Respiratory ratio,Peak airway pressure,Mean airway pressure,PEEP,Inspiratory plateau pressure,FiO2(optional)

Alarm Setting:

Tidal volume:High/Low;Minute ventilation:High/Low;Respiratory rate:High/Low;

FiO2(optional):High/Low;Airway pressure:High/Low ;FiCO2 (optional):High/Low

 

Continuous Positive Airway Pressure:Greater than (PEEP +15) cmH2O, 

15 s sustained when an alarm is triggered.

Negative pressure alarm:Airway pressure is less than (-10) cmH 2 O.

Apnea alarm:Setting time is 10 ~ 60 s

Gas deficiency alarm:less than 280 kPa.

AC power failure alarm:The mains breaks down or the power cord disconnects

Low battery alarm: Time is 20 min.

Alarm battery is exhausted: time is 10 min.

Silent Alarm:≤ 120 s

O2 sensor failure:FIO2 <15% Vol;

Dolphinmed is one professonal producer of  ICU ventilator, anesthesia machine, medical air compressor, anesthesia vaporizer, anesthesia ventialtor, veterinary anesthesia machine,  etc.  Dolphinmed are committed to provide the more perfect products and service for clients, to be the most valued medical equipment enterprise. 